  • The unit should have separate generator for mono-polar and bi-polar.

  • Should be Compatible for both open and laparoscopic surgery.

  • Should have facility to connect two mono-polar active pencils with hand & foot operation.

  • Should have separate digital display of power settings for bipolar and mono-polar cut and coagulation modes.

  • Should have return electrode contact safety.

  • Should have maximum range mono-polar cut power of at least 300 Watts variable in steps of 1 watts from 0 to 400 max with a ramp of slow moving inetially then fast for reaching at safety point.